Part of the project Addressing Biases in Neurological and Respiratory Monitoring Technologies

A pulse oximeter clipped to your finger says 94%. The doctor relaxes. But the oxygen in your arteries may actually be below 88%, low enough to need treatment.

That gap has a name: occult hypoxaemia. In 2020, a team at the University of Michigan compared more than 10,000 pairs of finger-clip readings with arterial blood tests. In their multi-hospital cohort, the hidden low-oxygen reading turned up in 17% of measurements from Black patients, compared with 6% from White patients (Sjoding et al., NEJM 2020). Same device, same reading, very different truth.

I work on this problem at the AI Healthcare Research Lab at Carnegie Mellon University Africa, on a study of bias in neurological and respiratory monitoring for African populations, under the guidance of Dr. Carine Mukamakuza and Prof. Sossena Wood. This post is the first in a short series about what we’re learning from the literature and the field.

Two devices, one pattern

Pulse oximeters shine red and infrared light through your skin and estimate oxygen from how much is absorbed. Melanin also absorbs light, and it does so differently at each wavelength. The UK’s 2024 independent review of medical devices put the root cause plainly: the light reacts differently with different levels of melanin, and the problem was made worse by testing devices on people with light skin and treating those readings as the norm (Equity in Medical Devices review). The review found extensive evidence that oximeters over-estimate oxygen in people with darker skin.

Diagram comparing light passing through lighter and darker skin to a pulse oximeter sensor

The same physics affects other light-based tools. Prof. Wood’s group at Carnegie Mellon measured skin melanin with a colorimeter and showed that it significantly affects signal quality and measured oxygen saturation in near-infrared spectroscopy (NIRS), a close cousin of the pulse oximeter used to monitor oxygen in the brain (Roy et al., 2024). Related work from the same Carnegie Mellon collaborators shows how the bias in pulse-oximetry algorithms grows as signal quality falls (Cao et al.).

EEG records the brain’s electrical activity through electrodes that must touch the scalp. Most were designed for straight, fine hair. Coarse, curly hair pushes electrodes away and raises the electrical resistance at the contact, which degrades the signal. The usual response in research has been quieter, and worse: people with that hair are simply left out. One review of 81 EEG papers in affective neuroscience found only five that explicitly reported Black American participants (position paper in Affective Science).

The pattern is the same in both cases. A device is designed and validated on a narrow population. It works well enough there to become the standard. Then it is deployed everywhere, and the error lands on the people who were never in the test group.

Why this matters more in Africa

In much of Africa, these are not niche tools.

Pneumonia remains the leading cause of death in children under five in sub-Saharan Africa (recent trial report). The pulse oximeter is one of the cheapest, most powerful tools for deciding which child needs oxygen or urgent referral. One estimate cited in the literature suggests that adding oximetry to standard childhood-illness guidelines could prevent about 150,000 deaths a year across Africa and Asia (summarised here).

So the device whose accuracy is weakest for darker skin is being scaled up fastest in places where most patients have darker skin. EEG faces the same mismatch: it is how clinicians diagnose and monitor epilepsy, brain injury and sleep disorders. A cap that can’t make contact through the most common hair types on the continent is a cap that leaves patients undiagnosed.

And the evidence base is thin exactly where it matters. Much of the clinical evidence on harm so far comes from US hospitals. We know far less about how these devices perform in African clinics, on African patients, under African conditions: heat, dust, limited power, and very sick children.

What can fix it

The good news is that this is an engineering and validation problem, and the field is moving. The fixes fall into four groups.

  1. Better hardware. Researchers at Carnegie Mellon developed Sevo electrodes, which work with cornrowed hair rather than against it. Braiding exposes the scalp along standard electrode positions, and a clip uses the hair itself to hold the electrode in place. In early tests this cut impedance by about 10 times compared with standard gold-cup electrodes (Etienne et al., 2020). Jasmine Kwasa and colleagues have since evaluated Sevo in clinical settings (Kwasa et al., 2021), and as a Fulbright scholar in Kenya she worked to expand access to quality epilepsy monitoring there (CMU). On the optical side, researchers are studying light sources and sensor layouts that are less affected by melanin.
  2. Better calibration and algorithms. If one calibration curve can’t fit every skin tone, models can account for skin pigmentation directly, using physics-informed or machine-learning approaches. That only works if they’re trained and tested on the right data, starting with measuring skin tone objectively, with a colorimeter rather than by eye, as the Wood lab recommends.
  3. Better testing rules. In January 2025 the US FDA proposed new guidance asking manufacturers to test pulse oximeters on more participants across the full range of skin pigmentation (FDA). The UK review made 18 recommendations of its own.
  4. Testing where devices are used. In February 2026, the Uganda Hypoxia Lab opened as the first independent device-testing facility of its kind in sub-Saharan Africa. That is the direction we need: evidence generated on the continent, by people from the continent.

Until better devices are everywhere, there are practical steps too. Clinicians can treat a borderline reading as a prompt for a closer look, not a final answer. Researchers can measure and report skin tone and hair type, and stop excluding participants by default.

What comes next

Our team is preparing a review of what is known about bias across these monitoring technologies: where the evidence is strong, where it is missing, and what is most worth building. I’ll share it here when it’s ready, along with shorter posts on each device.

If you design medical devices, run a clinic or a lab, fund health technology, or simply have a story about a reading that didn’t add up, I’d like to hear from you. The fix will come from engineers, clinicians and patients working on it together, and much of that work should happen here, in Africa.

Isaac Thani is the founder of Kainetics Labs and a Research Associate at the AI Healthcare Research Lab, Carnegie Mellon University Africa, in Kigali.


Views here are my own and do not represent Carnegie Mellon University Africa, my collaborators or funders. Posts on health topics are for general information, not medical advice.

Cite this post

Thani, I. (2026, November 23). When medical sensors don't see everyone. Isaac Thani. https://zikt.github.io/research/2026/11/23/when-medical-sensors-dont-see-everyone/